Four versions of the closed cup flashpoint testers are available. All functions are accessed by a two-button keypad and the SMART control feature (Seta Multifunctional And Rotational Test). The digital display shows test parameters, instrument status and the test results. Audible prompts are sounded when an action is required by the user. The automatic flash detector reduces the chance of misinterpretation of the results and enhances repeatability. Instruments have a integral gas tank which is filled using a standard butane (lighter) refill cartridge and are suitable for mobile use when operated with the optional 12 volt power supply adapter ( please ask for details).
- Flash point tests in less than two minutes
- Small sample sizes of 2 or 4 ml
- Audible and on screen prompts for ease of use
Closed cup tester: As above with the addition of a temperature ramp mode that allow the rapid determination of the flash point of a sample in accordance with ASTM D7236 and IP 534.
Active cool closed cup tester: Active Cool function to rapidly and efficiently carry out Flash/No flash tests and extend the temperature down to 10 °C - Cooling and heating with Peltier cells.
Active cool corrosion resisting closed cup tester: Flash/No Flash testing (10 to 130 °C) rapid temperature normalisation between tests. Corrosion resisting insert in the sample cup.
Models with ramp mode. As above plus ASTM D7236 and IP 534.
- Catalog No:
- 751-0066
- 751-0065
- 751-0013
- Model:
- 33200-2
- 33000-0
- 31000-0
- Temperature range:
- 10…130 °C
- Ambient…300 °C
- Ambient…300 °C
- Test modes:
- Rapid equilibrium and ramp 2 °C/minute
- Rapid equilibrium and ramp 2 °C/minute
- Rapid equilibrium and ramp 2 °C/minute
- Cup material:
- Aluminium
- Aluminium
- Blackened aluminium
- Sample size:
- 2 ml for flash points up to 100 °C; 4 ml for flash points from 100 to 300 °C
- 2 ml for flash points up to 100 °C; 4 ml for flash points from 100 to 300 °C
- 2 ml for flash points up to 100 °C; 4 ml for flash points from 100 to 300 °C
- Test time:
- 1 to 99 min.
- Test duration RE mode:
- 1 min for flash points up to 100 °C, 2 min for flash points 100 to 300 °C or user determined 1 - 99 min
- 1 min for flash points up to 100 °C, 2 min for flash points 100 to 300 °C or user determined 1 - 99 min
- 1 min for flash points up to 100 °C, 2 min for flash points 100 to 300 °C or user determined 1 to 99 min
- Power:
- 200 W
- 200 W
- 200 W
- W×D×H:
- 280×260×260 mm
- 280×260×260 mm
- 280×256×256
- Weight:
- 5 kg
- 4 kg
- 4 kg