This rugged, waterproof portable meter is designed for a wide range of pH, ion concentration, conductivity, dissolved oxygen and temperature testing and field applications. It can be used in the most demanding locations thanks to its IP67-rated housing. The meter offers three measuring channels allowing users to measure pH, conductivity and DO simultaneously or view each channel separately.
- Up to five point pH calibrations with calibration editing option that allows individual points to be fixed without a full recalibration
- Automatically compensate dissolved oxygen readings using ATC temperature, manual salinity input and built-in barometer for barometric pressure readings
- Informative, easy-to-read backlit graphic display with clear onscreen instructions simplifies operation
- Measurement stability indicator and selectable read modes – Auto-Read holds a stable reading, Continuous shows changing readings with hold option and Timed logs data at specific intervals
- Data log collects and date/time stamps up to 5000 measurement sets with optional sample ID and user ID
- Calibration log stores ten most recent calibrations
- Orion Star Com software facilitates data transfer to a computer via USB or RS232 connections
Measure pH, mV, relative mV or ORP with temperature on channel one and measure dissolved oxygen in percent saturation or mg/L with temperature using RDO optical DO probes or polarographic DO probes on channel two.
Automatically compensate dissolved oxygen readings using ATC temperature, manual salinity input and built-in barometer for barometric pressure readings. Transfer single, range or all data points to printer or computer.
Auto-ranging conductivity values offer consistent readings across entire measurement range without need to recalibrate. Use linear, non linear, EP (USP) or off conductivity temperature compensation with 5, 10, 15, 20 or 25 °C reference temperatures. 3-years replacement warranty.
The kit additionally includes: 8107UWMMD ROSS Ultra Triode gel filled epoxy-body pH/ATC electrode (3 m cable), 013010MD DuraProbe 4-cell (K=0,475) epoxy-body conductivity sensor (3 m cable), 087010MD RDO optical DO sensor (3 m cable), calibration sleeve and guard, Orion pH 4/7/10 buffer pouches (10 packs each), ROSS electrode storage solution (475 ml), Orion rinse solution pouches (10 packs), Orion 1413 µS conductivity standard pouches (10 packs), Orion Star A series portable meter armor and Orion Star A series hard-sided field case.
- Catalog No:
- 665-0282
- 665-0281
- Model:
- A329
- A329
- pH range:
- –2,000...+20,000
- –2,000...+20,000
- pH resolution:
- 0,1 / 0,01 / 0,001
- 0,1 / 0,01 / 0,001
- pH accuracy:
- ±0,002
- ±0,002
- pH calibration:
- Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
- Up to 5 points; Automatic buffer recognition (USA/NIST and DIN)
- mV range:
- ±2000,0
- ±2000,0
- mV resolution:
- 0,1
- 0,1
- mV accuracy:
- ±0,2 mV or 0,05%
- ±0,2 mV or 0,05%
- ISE range:
- 0,0001 - 19900
- 0,0001 - 19900
- ISE resolution:
- 0,0001 minimum, 1 to 3 significant digits (user selectable)
- 0,0001 minimum, 1 to 3 significant digits (user selectable)
- ISE accuracy:
- ±0,2 mV or ±0,05% of reading, whichever is greater
- ±0,2 mV or ±0,05% of reading, whichever is greater
- ISE calibration:
- Up to 5 points
- Up to 5 points
- ISE units:
- ppm, M, mg/l, %, ppb, none
- ppm, M, mg/l, %, ppb, none
- Conductivity range:
- 0,001 µS/cm - 3000 mS/cm
- 0,001 µS/cm - 3000 mS/cm
- Conductivity resolution:
- 0,001 μS minimum, auto ranging, up to 4 significant digits
- 0,001 μS minimum, auto ranging, up to 4 significant digits
- Conductivity accuracy:
- ±0,5% of reading ±1 digit
- ±0,5% of reading ±1 digit
- Conductivity cell constant:
- 0,001 - 199,9 cm⁻¹
- 0,001 - 199,9 cm⁻¹
- Conductivity calibration:
- Up to 5 points
- Up to 5 points
- Conductivity reference temperature:
- 5/10/15/20/25 °C
- 5/10/15/20/25 °C
- Resistivity range:
- 2 Ω/cm - 100 MΩ/cm
- 2 Ω/cm - 100 MΩ/cm
- Resistivity accuracy:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- Resistivity resolution:
- 1 ohm or 0,1 meg-ohm, auto ranging
- 1 ohm or 0,1 meg-ohm, auto ranging
- Salinity range:
- 0,06 - 80,00 psu / 0,05 - 42,00 ppt
- 0,06 - 80,00 psu / 0,05 - 42,00 ppt
- Salinity resolution:
- 0,01 psu or 0,01 ppt minimum, auto ranging
- 0,01 psu or 0,01 ppt minimum, auto ranging
- Salinity accuracy:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- TDS range:
- 0,001 - 200,0 ppm
- 0,001 - 200,0 ppm
- TDS resolution:
- 0,001 ppm minimum, auto ranging, up to 4 significant digits
- 0,001 ppm minimum, auto ranging, up to 4 significant digits
- TDS accuracy:
- 0,5% of reading ±1 digit
- 0,5% of reading ±1 digit
- TDS conversion factor:
- Linear (0,02 - 9,99)
- Linear (0,02 - 9,99)
- DO concentration:
- 0 - 90 mg/L (polarographic); 0 - 50 mg/L (RDO)
- 0 - 90 mg/L (polarographic); 0 - 50 mg/L (RDO)
- DO concentration resolution:
- 0,1 / 0,01 mg/L
- 0,1 / 0,01 mg/L
- DO saturation:
- 0 - 600 % (polarographic); 0 - 500 % (RDO)
- 0 - 600 % (polarographic); 0 - 500 % (RDO)
- DO accuracy:
- Polarographic: ±0,2 mg/l or ±2% of reading, whichever is greater
RDO: ±0,1 mg/l (up to 8 mg/), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l) - Polarographic: ±0,2 mg/l or ±2% of reading, whichever is greater
RDO: ±0,1 mg/l (up to 8 mg/), ±0,2 mg/l (8 - 20 mg/l), ±10% of reading (up to 50 mg/l)
- DO calibration:
- Water-saturated air, air saturated water, manual (Winkler) and zero point
- Water-saturated air, air saturated water, manual (Winkler) and zero point
- DO barometric pressure range:
- Automatic or manual, 450 - 850 mmHg
- Automatic or manual, 450 - 850 mmHg
- Temperature range:
- –5,0...+105,0 °C (–5,0...+50,0 °C with DO)
- –5,0...+105,0 °C (–5,0...+50,0 °C with DO)
- Temperature resolution:
- 0,1 °C
- 0,1 °C
- Temperature accuracy:
- ±0,1 °C
- ±0,1 °C
- Temperature calibration:
- 1 point temperature offset calibration
- 1 point temperature offset calibration
- Temperature compensation:
- Automatic or manual
- Automatic or manual
- Pressure range:
- 400 - 850 mm Hg
- 400 - 850 mm Hg
- Pressure resolution:
- 1 mm Hg
- 1 mm Hg
- Display:
- Graphic LCD with backlight
- Graphic LCD with backlight
- Battery type:
- 4 AA (included)
- 4 AA (included)
- Battery life:
- 800 hours
- 800 hours
- Power supply:
- AC adapter or 4 AA batteries
- AC adapter or 4 AA batteries
- Compliance:
- IP 67
- IP 67
- Data logging - log function:
- Manual / Automatic with AUTO-READ™
- Manual / Automatic with AUTO-READ™
- Data logging capacity:
- 5000
- 5000
- Memory:
- Non-volatile memory preserves data log, calibration log and meter settings
- Non-volatile memory preserves data log, calibration log and meter settings
- Languages:
- English, Spanish, French, Italian, German and Chinese; Portuguese and Korean available with software update
- English, Spanish, French, Italian, German and Chinese; Portuguese and Korean available with software update
- IP rating:
- IP 67
- IP 67
- Connections:
- BNC, pin tip, 8- and 9-pin mini-DIN
- BNC, pin tip, 8- and 9-pin mini-DIN
- Outputs:
- RS232, USB
- RS232, USB
- Input:
- 3 measuring channels
- 3 measuring channels
- Alarm:
- High/low limit alarm, calibration due alarm
- High/low limit alarm, calibration due alarm
- No. of channels:
- 3
- 3
- Calibration:
- pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ
- pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ
- Calibration points:
- 1 point relative mV (RmV) mode, 1 point ORP mode
- 1 point relative mV (RmV) mode, 1 point ORP mode
- W×D×H:
- 60×105×240 mm
- 60×105×240 mm
- Weight:
- 0,5 kg
- 0,5 kg